The Bus page lets you send raw transactions on the station’s I²C, SPI, UART and socket channels. You can type the bytes to send in hex or as text, view what comes back in a hex dump, read it as numbers, and keep a history of what you ran.
Accessing
In the side menu select Bus (always present; a bus type you have no channel for is disabled). Firmware 6.0 or later is needed.
Transactions go straight to the bus — nothing checks what a device does with the bytes. A wrong write can reconfigure or damage connected hardware, and a long receive holds the bus for every other client. The first time in a browser session you are asked to confirm.
Choose the bus, channel and action
Select the bus type (I²C, SPI, UART, Socket), the Channel (its own settings — clock speed, SPI mode, baud rate, bus type, termination byte — are filled in for you, and can be changed for this transaction), and the Action:
- I²C — Scan, Send, Receive, Send and receive.
- SPI — Send and receive, Send, Receive.
- UART — Send and receive, Send, Receive, Break, Clear buffers.
- Socket — Send and receive, Send, Receive.
Bus settings
- I²C — 7-bit Address (01–7F); Scan lists the addresses that answer and offers a Found list to pick from.
- SPI — Mode0–Mode3, clock speed, chip-enable-active-high.
- UART — baud rate (9,600–921,600), parity, flow control, bus type (RS232/RS422/RS485).
- Socket — host and port.
- Timeout / termination (UART and Socket) — receive timeout (max 15,000 ms) and a termination byte (default
0A). - Bytes to receive (1–65,536) for Receive and Send-and-receive; with a termination byte this is the maximum.
Send and view data
Type the bytes in the Send box, choosing Hex (AA BB 01) or Text (*IDN?\n, with \r \n \t \0 \\ \xNN). The page shows the byte count and the same bytes in the other form. Load fills the box from a file (up to 64 kB). Click the run button to execute.
The Received panel shows the byte count and a hex view (offset, up to 16 bytes, and the characters). Click a byte to select it as the start for the number readouts, which show it as 8/16/32/64-bit signed and unsigned integers and 32/64-bit floats, in little- or big-endian. Hex copies the received bytes; Save writes them as .bin, .txt or a .csv table.
History and saved transactions
The History tab lists this session’s transactions (newest first) with time, bus, channel, action, bytes and duration; you can Run again, Load into the form, export CSV or Clear it (history is not kept across reloads). Save a transaction with Save…; the Saved tab lets you run, load or delete saved transactions.