# Web UI – Audio analyser

> Turns a Tone Audio Module (ESH10000560) into a two-channel audio generator and analyser in the browser — level, frequency, THD+N, SNR, spectrum and scope.

_Source: E-Sharp (esharp.se) · Accordion Web UI_


The Audio analyser page turns a Tone Audio Module (ESH10000560) into a two-channel audio generator and analyser in your browser. You set a test signal on each output, choose which input each channel measures, and read level, frequency, THD+N and SNR live, together with a spectrum, a waterfall and a scope view. The module does the signal processing itself at 96 kSps and 24 bit — only settings and results travel between the module and your browser.

## Accessing

Open the station in a browser (firmware 6.5 or later), then **Instrument Panel → Audio analyser**. The page appears when the station has a Tone Audio Module; each module has its own panel titled with its slot and name. The **Analyser running** switch starts and stops the measurements.

## Channel readings

Each channel panel (CH1 left, CH2 right) summarises its settings and shows: **Vrms** and **Vpp** (input level at the pin), **Frequency** (strongest tone), **THD+N** (%), **SNR** (dB) and **Clip** (red when the input reaches the top of the range — the other readings are unreliable while it is red). The **Output** switch turns the channel's output jack on or off; the generator keeps running while the output is off. The readings are also station channels (e.g. `0.2.ESH10000560.MEAS1_THD`), so you can add them to the Chart Viewer.

## Generator and input settings

**Generator** (per channel): On, **Waveform** (Sine, Square, Triangle, White noise, Pink noise, Sweep), **Frequency** (Hz), **Duty** (% for square), **Level** (% of full output), and **Input to output**. At 100 % level a sine is about 1.4 Vpp (0.49 Vrms). **Input**: which input pair (IN1–IN4) the channel measures, and **Gain** in dB (−6 dB normal; gain acts on IN1 and IN4 only). **Measurement filters**: a first-order **low-pass** (off, 24/20/16/10/5/2/1 kHz) and **Weighting** (Flat or A-weighted, applied to the spectrum, Vrms, THD+N and SNR).

## Analysis view

The toolbar: **Live** (freeze/continue), **View** (Spectrum or Waterfall), per-channel show/hide, **Averaging** (off, 2/4/8 frames), **Slope**, **Peak**, **Frequency axis** (log/linear), **Level unit**, **Harmonics** (marks f0 and H2–H5 with levels), **Hold reference**, **Reset zoom**, and **CSV**/**PNG** export. **Peak** draws the highest level at each frequency (last 3/5/10 frames, **Max hold**, or **Sweep response**). The **Waterfall** view shows the spectrum over time, newest at the top, colour by level.

## Scope and output waveforms

Turn on **Scope** to see the measured input signal in time (with DC and AC-RMS readouts) next to the signal the outputs are set to produce. **Scope trigger** (e.g. CH1 rising) steadies a repeating signal.

## Measurement range and limits

Measurements are valid 50 Hz–20 kHz. All inputs and outputs are line level: on IN1/IN4 keep the input at or below 0.7 Vrms (0.85 Vrms max) at −6 dB gain; on IN2/IN3 at or below 0.35 Vrms (0.50 Vrms max). A square wave reads a high THD+N (≈47 %) by nature, and its edges ring slightly (≈9 % overshoot) from the converter's filter — a square wave above ≈91 % of full scale can turn Clip red for the same reason. Terminate unused inputs when measuring low levels.